next up previous
Next: Anode Calibration Up: PHENIX Muon Tracking Calibration Previous: Fine Cathode Plane

Coarse Cathode Calibration

The coarse cathode planes will be read out with a latch so they will not require as good of a calibration as the fine cathode planes. However, the threshold levels should be set uniformly across a given chamber, so the strips should be pulsed at a given voltage with several threshold settings to calibrate the system. Either the strips can be pulsed directly, or the front end electronics can be pulsed. There are approximately 8000 coarse cathode channels, and we would like to pulse the system at one voltage and five threshold settings, with 300 data points at each threshold setting. This gives a total of 1.2x10 data points required for calibration. In addition, we would like to have the capability to pulse the coarse cathodes at different voltage levels for test purposes, but will probably only need a single voltage level for a standard calibration run.

The data stored for each coarse cathode strip is the threshold value at which the channels fire 50% of the time (8000 calibration constants). The calibration should be performed approximately monthly or on the same schedule as the fine cathode calibration.



Converted to HTML from LaTeX by Wayne Kinnison
Fri Aug 25 17:58:27 MDT 1995