next up previous
Next: Coarse Cathode Calibration Up: PHENIX Muon Tracking Calibration Previous: Introduction

Fine Cathode Plane Calibration

The ADC calibration of all strips relative to each other in a given chamber will need to be known to less than 1% in order to obtain a 100 m resolution from the fine cathode planes, if the readout pitch is 1 cm. To achieve this, all strips will have to be pulsed with a common pulser over several pulseheights spanning the range of the ADC. The strips in a given chamber may be pulsed in sets of every fourth (or more) strips at a time to avoid cross talk. If a calibration error less than 1% is to be achieved, a precision pulser must be used, and the capacitive coupling of the pulse to the strips must be such that the error in the known capacitance is less than 1%. This can be achieved either by using precision capacitors which are uniform to the sub-1% level or by using capacitive strips whose area is known to less than 1% and whose distance from the strips is constant.

The North muon arm consists of approximately 16000 channels of fine cathode strips, and we would like to take calibration data for approximately 30 different voltage levels and on the order of 300 data points at each voltage. This gives a total of 1.44x10 data points required for ADC gain calibration. Twice this number of data points will be required for both the North and South muon arms to be calibrated.

In addition to the ADC gain calibration, the pedestal levels of all fine cathode channels must be known. To do this we would like to take approximately 1000 data points at zero voltage. This same measurement will allow us to measure the noise levels of all channels, which is a critical measure of how well the system is performing. To do the pedestal calibration a total of 1.6x10 data points will be needed.

The data stored for each fine cathode strip will be: 1) the fit parameters for the ADC gain calibration (one parameter if it is very linear, more likely a few parameters) 2) the pedestal values and 3) the sigmas of the pedestal values. This adds up to approximately 80,000 calibration constants for the fine cathode planes.

The calibrations should probably be performed bi-monthly at a minimum since the performance of the chambers depends so strongly on a good fine cathode calibration and low noise levels (which would be checked with a calibration). It would also be useful to track a small fraction of the channels on a short time scale in between calibrations to make sure that calibrations are not drifting significantly. This would be especially useful if the full calibration can only be performed infrequently.



next up previous
Next: Coarse Cathode Calibration Up: PHENIX Muon Tracking Calibration Previous: Introduction



Converted to HTML from LaTeX by Wayne Kinnison
Fri Aug 25 17:58:27 MDT 1995