Si + MCM Assembly QA with an LED


Look here for some history of this QA procedure.

Look here for a few details on the setup.

The QA procedure with the LED

  1. Carefully put the assembly on the detector mount on the platform of the x-y table and connect the MCM output cable to the J6 of the power/comm board. Be sure to wear a ground strap and use a vacuum pen to pick up the Si-strip part.

  2. Slide in the conductive rubber under the small (bias) pad on the kapton cable from the Si to the MCM, on the left side. Gently squeeze the kapton cable and the conductive rubber against the copper surface of the detector mount with a small clamp. Screw down the Si detector on the detector mount. The result is a Si-bias connection through the conductive rubber and the surface of detector mount.

  3. Turn on the stepper motor controller and position the LED above the center of the Si detector (as close as you can).

  4. Shield the whole setup by carefully covering the assembly with a black cloth or two and closing the top of the steel box.

  5. Turn on everything.

  6. Start MCM control software (pcb_mcm, mcm_read, mvd_main) on mvdonl5 (a Pentium-PC for the assembly QA). The easiest way to do this is to start mvd_main and then run the "control.txt" script. Look here for details if you are not sure how to do this.

  7. In the Mcm_read control panel, click on "LED Test". This pops up the the LED test control panel. Then do the following:

  8. There should be a file called c:\mvd\histos\led_[assembly name].dat created, where [assembly name] is something like OB-04.

  9. Power down before you open up the box: the 4 Lambda units, the Keithly-487, the supply under the laser station.

  10. If you want to move the xy table, the buttons on the front panel are the easiest way when the light box is open. If you want to adjust the positions of the LEDs during the tests, look here for some instructions.


    last update: 4-Jan-2001
    John Sullivan (sullivan@lanl.gov) & HvH