Humidity sensor tests


Omega HX11V:
    [5" × 5/8" cylinder, 4-wire output]
    supply 12V
    output 1-5V for 0%-100%
    %RH = (V-1)*25
              Vout  (V-1)*25%       Fisher           where
            -----------------------------------------------------
    tests:  1.08V      2.0%            -        Sanpia Dry Keeper
            1.43V     10.8%           23%       ambient
            5.00V    100  %       after cal.    saturated breath
    
Omega HX49-D-V
    [6" × 1/2" cylinder on 2"×2.5"×4.5" box]
    supply 12V
    output 0-5V for 0%-100%
    %RH = V*20
            Vout   V*20%   Fisher    Where
           --------------------------------
    tests: 0.006    0.1%    20%      Sanpia Dry Keeper
           0.56V   11.2%    22%      ambient
           4.56V   91.2%     -       saturated breath

    Second unit: same results.

  
Omega RHCN-1C
    [big plastic gray threaded cylinder, plus control box]
    no manual

    





Last update 9 Mar 2000 - HvH