TITLE>MuTr Mechanics Fault Chart

Muon Tracker Mechanics Fault Chart

CSC Gas System

Fault

System Effect

Possible Consequence

Monitors

Response

Chamber Leak

Mass flow difference between input and output

Increase in O2 and H2O levels in gas

Change in isobutane/CF4 ratio in gas

Change in difference pressure across chamber

Chamber currents increase

Chamber HV trips

Flammable gas in magnet volume

Fire

Loss of chamber efficiency

Flammable gas monitors

Gas composition monitors

Mass flow meters

Turn off pump

Close chamber valves or Nitrogen purge

Turn off HV

Initiate operator alarm

Over pressure

Differential pressure to atmosphere at chamber

Blow window or distort panel

Differential pressure monitor to atmosphere

Passive bubbler pressure relief

Close backflow valve at pump

Inform operator

Under pressure

Differential pressure to atmosphere at chamber

Implode window or distort panel

Differential pressure monitor to atmosphere

Passive bubbler pressure relief

Open backflow valve at pump

Inform operator

Gas purity

Increase in O2 and H2O levels in gas

Change in isobutane/CF4 ratio in gas

Chamber currents increase

Chamber HV trips

Loss in chamber efficiency

O2 and H2O ppm monitors

Isobutane/CF4 ratio monitor

Inform operator

Monitoring equipment power failure

(Gas hut)

Overall gas flow instantly reads zero

All pressures in gas hut read zero

Gas purity monitors read zero

Chamber currents increase

Chamber HV trips

ONCS power monitor shows trip

Shut off pump

Turn off HV

Initiate operator alarm

Monitoring equipment power failure (Magnet)

Chamber input and output reads zero

All pressures at manifolds read zero

Chamber currents increase

Chamber HV trips

ONCS power monitor shows trip

Shut off pump

Turn off HV

Initiate operator alarm

Solenoid valves power failure

none

Loss of control

ONCS power monitor shows trip

Alarm operator

Pump power failure

Overall gas flow goes to zero

Differential pressure across pump goes to zero

Chamber currents increase

Chamber HV trips

Differential pressure monitor

Absolute pressure monitors

Turn off HV

Initiate operator alarm

Fire

Heat and smoke

Loss of chamber and electronics

Fire alarm

Smoke detector

Shut off pump

Close chamber valves or nitrogen purge

Turn off HV

Initiate operator alarm

Detector Systems

Fault

System Effect

Possible Consequence

Monitors

Response

High leakage current

HV channel senses increase in current

Broken wire

spark

HV module

HV channel trips off

Initiate operator alarm

Broken Wires

HV short

Damage to cathode foil

Spark

HV module

HV channel trips off

Initiate operator alarm

High Voltage Systems

Fault

System Effect

Possible Consequence

Monitors

Response

High current draw on individual channel

HV channel senses increase in current

Broken wire

Spark

HV module

HV channel trips off

Initiate operator alarm

Requested and readback voltage error

Incorrect voltage

Inoperative chamber

Loss of chamber efficiency

ONCS

HV channel trips off

Initiate operator alarm

DC power failure in crate

Bad readback or current or voltage value

Loss of chamber voltage

Damage to wire

ONCS

Turn off HV and crate

Initiate operator alarm

AC power failure

All channels voltages and currents go to zero

Loss of Efficiency

ONCS

Initiate operator alarm

HV is off


Edited by: Dave Lee, Los Alamos National Laboratory, Los Alamos, NM