TITLE>MuTr Mechanics Fault Chart
Fault |
System Effect |
Possible Consequence |
Monitors |
Response |
---|---|---|---|---|
Chamber Leak |
Mass flow difference between input and output Increase in O2 and H2O levels in gas Change in isobutane/CF4 ratio in gas Change in difference pressure across chamber |
Chamber currents increase Chamber HV trips Flammable gas in magnet volume Fire Loss of chamber efficiency |
Flammable gas monitors Gas composition monitors Mass flow meters |
Turn off pump Close chamber valves or Nitrogen purge Turn off HV Initiate operator alarm |
Over pressure |
Differential pressure to atmosphere at chamber |
Blow window or distort panel |
Differential pressure monitor to atmosphere |
Passive bubbler pressure relief Close backflow valve at pump Inform operator |
Under pressure |
Differential pressure to atmosphere at chamber |
Implode window or distort panel |
Differential pressure monitor to atmosphere |
Passive bubbler pressure relief Open backflow valve at pump Inform operator |
Gas purity |
Increase in O2 and H2O levels in gas Change in isobutane/CF4 ratio in gas |
Chamber currents increase Chamber HV trips Loss in chamber efficiency |
O2 and H2O ppm monitors Isobutane/CF4 ratio monitor |
Inform operator |
Monitoring equipment power failure (Gas hut) |
Overall gas flow instantly reads zero All pressures in gas hut read zero Gas purity monitors read zero |
Chamber currents increase Chamber HV trips |
ONCS power monitor shows trip |
Shut off pump Turn off HV Initiate operator alarm |
Monitoring equipment power failure (Magnet) |
Chamber input and output reads zero All pressures at manifolds read zero |
Chamber currents increase Chamber HV trips |
ONCS power monitor shows trip |
Shut off pump Turn off HV Initiate operator alarm |
Solenoid valves power failure |
none |
Loss of control |
ONCS power monitor shows trip |
Alarm operator |
Pump power failure |
Overall gas flow goes to zero Differential pressure across pump goes to zero |
Chamber currents increase Chamber HV trips |
Differential pressure monitor Absolute pressure monitors |
Turn off HV Initiate operator alarm |
Fire |
Heat and smoke |
Loss of chamber and electronics |
Fire alarm Smoke detector |
Shut off pump Close chamber valves or nitrogen purge Turn off HV Initiate operator alarm |
Fault |
System Effect |
Possible Consequence |
Monitors |
Response |
---|---|---|---|---|
High leakage current |
HV channel senses increase in current |
Broken wire spark |
HV module |
HV channel trips off Initiate operator alarm |
Broken Wires |
HV short |
Damage to cathode foil Spark |
HV module |
HV channel trips off Initiate operator alarm |
Fault |
System Effect |
Possible Consequence |
Monitors |
Response |
---|---|---|---|---|
High current draw on individual channel |
HV channel senses increase in current |
Broken wire Spark |
HV module |
HV channel trips off Initiate operator alarm |
Requested and readback voltage error |
Incorrect voltage Inoperative chamber |
Loss of chamber efficiency |
ONCS |
HV channel trips off Initiate operator alarm |
DC power failure in crate |
Bad readback or current or voltage value |
Loss of chamber voltage Damage to wire |
ONCS |
Turn off HV and crate Initiate operator alarm |
AC power failure |
All channels voltages and currents go to zero |
Loss of Efficiency |
ONCS |
Initiate operator alarm HV is off |