These tables are for sensor failure only. They assume that the safety/interlocks presented in the general fault charts are in operation.
Unified Air Cooling System Sensors |
|||
Failing sensor |
Capabilities of that sensor |
Means of diagnosing or double checking |
Consequences of failure |
The humidity sensor | Talk only--no automated shutdown under any circumstances | No direct double check; there is only one humidity sensor. The time between signals from the liquid level sensor is humidity-related and may help. | We would have to run system uncontrolled or go in to fix the sensor. High humidity could affect the placement of the detectors, requiring changes in vertex finding algorithms. |
A flowmeter (one of 3) | Hardware shutdown if 2 out of 3 report trouble | There are 3 flowmeters such that F1 = F2 + F3. | If one fails, we can fill in its measurements using easy math. If another also fails, we can't. |
One MCM temp sensor fails | Eventual software shutdown (algorithm undetermined--probably shut down in blocks of six MCMs) | Can check MCM above/below that one in flowstream; if there is a discrepancy, ignore the failing temp sensor from then on. Also can check signal to noise ratio for the affected MCM. | One temp sensor doesn't work and can't be used for measurements or to double check another sensor. |
Liquid level sensor | Controls a valve that drains water out of loop | If average time between signals is recorded, can sense deviation from normal schedule. Can also keep draining system at intervals in the absence of signals from the sensor. | The draining of water from the dehumidifier will be less controlled. |
Temp sensors in the flow loop | Hardware shutdown if 2 out of 3 report trouble | Temp sensors in the loop should have some correlation between readouts. | Any one failure is okay, but system has one less doublecheck. |
FC-75 liquid cooling system sensors |
|||
Failing sensor |
Capabilities of that sensor |
Means of diagnosing or double checking |
Consequences of failure |
Temp sensor in flow loop (one of 2) | Hardware shutdown with flow sensor (F7 and T8, F8 and T10)
Hardware shutdown if both temp sensors report trouble |
Temp sensors on motherboard, and requiring both to show trouble before shutdown. Can still control with only one working. | Less redundancy in liquid system control; if T8 or T10 goes town, the flowmeter will be able to act alone. |
Flowmeter | Hardware shutdown with temp sensor (F7 and T8, F8 and T10) | No direct backup; only one flowmeter. Temp sensors double check to some degree. | Pump would have to run uncontrolled, and system could not directly detect leaks in the liquid loop |
Motherboard temp sensors (one of 3) | Software shutdown (algorithm undetermined) | As with MCMs, make sure they agree | As with MCMs, there's one fewer temp sensor for measurement and double checking. |
This page last modified on 7 September '98 by Rachel Cunningham.